Shopping Cart

No products in the cart.

AAMI ST55 2016 R2023

$125.88

ANSI/AAMI ST55:2016/(R)2023: Table-top steam sterilizers

Published By Publication Date Number of Pages
AAMI 2016
Guaranteed Safe Checkout
Category:

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

This standard establishes minimum construction and performance requirements for small tabletop steam sterilizers that use saturated steam as the sterilizing agent and that have a volume less than or equal to 56.63 liters (2 cubic feet). This standard is the fourth edition of Table-top steam sterilizers, which was first published as an American National Standard in 1997 as ANSI/AAMI ST55:1997. In comparison to the second edition, which was approved in 2003, this new edition covers cassette sterilizers (which were excluded from the scope of previous editions), incorporates revisions of the methodology for testing the biological performance of table-top steam sterilizers with dental handpieces, and includes a requirement that certain sterilizers be tested for noncondensable gases.

PDF Catalog

PDF Pages PDF Title
1 ANSI/AAMI ST55:2016/(R)2023; Table-top steam sterilizers
4 AAMI Standard
Copyright information
5 Contents
6 Glossary of equivalent standards
7 Committee representation
9 Foreword
11 1 Scope
1.1 General
1.2 Inclusions
1.3 Exclusions
2 Normative references
12 3 Definitions, symbols, and abbreviations
13 Table 1—Conversions of saturated pressure units at sea level
4 Requirements
4.1 Labeling
4.1.1 Device markings
4.1.1.1 Identification
14 4.1.1.2 Safety
4.1.2 Information manual
4.1.3 Service manual
15 4.2 Sterilizer design, construction, components, and accessories
4.2.1 Pressure requirements
4.2.2 Pressure vessel certification
4.2.3 Electrical components
4.2.4 Corrosion resistance
4.2.4.1 Sterilizer surfaces
4.2.4.2 Loading accessories
4.2.5 Air filters
4.2.6 Water supply reservoir
4.3 Sterilizer safety
4.3.1 Interlock
16 4.3.2 Prevention of thermal hazards
4.3.3 Sterilizer controls for aborting cycles
4.4 Process monitoring and control devices
4.4.1 General
4.4.2 Chamber temperature
4.4.2.1 Temperature monitoring and recording
4.4.2.2 Positioning of temperature sensors
4.4.2.3 Accuracy of temperature measurement
4.4.2.4 Resolution of temperature measurement
17 4.4.3 Sterilizer temperature control
4.4.4 Sterilizer exposure timer
4.4.5 Pressure measurement
4.4.5.1 Chamber pressure indicator
4.4.5.2 Jacket pressure indicator
4.4.6 Cycle completion
4.4.7 Sterilization fault conditions
4.4.8 Cycle documentation
18 4.5 Biological performance of sterilizers
4.6 Mechanical air removal
4.6.1 Air removal (dynamic-air-removal sterilizers)
4.6.2 Air leaks (prevacuum sterilizers)
4.6.3 Air removal (gravity-displacement cycles)
4.7 Moisture retention
4.8 Sterilizer performance certification and recordkeeping
5 Tests
19 5.1 Labeling
5.2 Sterilizer design, construction, components, and accessories
5.2.1 Pressure requirements
5.2.2 Pressure vessel certification
5.2.3 Electrical components
5.2.4 Corrosion resistance
5.2.4.1 Sterilizer surfaces
5.2.4.2 Loading accessories
5.2.5 Air filters
5.2.6 Water supply reservoir
5.3 Sterilizer safety
5.3.1 Interlock
5.3.2 Prevention of thermal hazards
5.3.3 Sterilizer controls for aborting cycles
5.4 Process monitoring and control devices
5.4.2 Chamber temperature
5.4.2.1 Temperature monitoring and recording
20 5.4.2.2 Positioning of temperature sensors
5.4.2.3 Accuracy of temperature measurement
5.4.2.4 Resolution of temperature measurement
5.4.3 Sterilizer temperature control
5.4.4 Sterilizer exposure timer
5.4.5 Pressure measurement
5.4.5.1 Chamber pressure indicator
5.4.5.2 Jacket pressure indicator
5.4.6 Cycle completion
5.4.7 Sterilization fault conditions
5.4.8 Cycle documentation
5.5 Biological performance of sterilizers
5.5.1 General testing requirements
21 5.5.2 Biological performance with a textile PCD (BI test pack)
5.5.2.1 Construction of the textile PCD
22 Figure 1—Three-towel PCD (BI test pack) for textile load
23 Figure 2—Towel folding procedure
Figure 3—Completed wrapped textile PCD (BI test pack)
5.5.2.2 Placement of the textile PCD
24 5.5.2.3 Cycle operation
5.5.2.4 Incubation of BIs
5.5.2.5 Acceptance criteria
5.5.3 Biological performance with liquid loads (if applicable)
5.5.3.1 Test flasks
5.5.3.2 Placement of test flasks
5.5.3.3 Cycle operation
5.5.3.4 Incubation of biological indicators
5.5.3.5 Acceptance criteria
25 5.5.4 Biological performance with wrapped instrument PCD (BI test tray)
5.5.4.1 Construction of the wrapped instrument PCD
5.5.4.2 Placement of the wrapped instrument PCD
5.5.4.3 Cycle operation
5.5.4.4 Incubation of biological indicators
5.5.4.5 Acceptance criteria
5.5.5 Biological performance with dental handpieces
5.5.5.1 Handpiece selection
26 5.5.5.2 Inoculation of the handpiece turbine
5.5.5.4 Handpiece placement during test
5.5.5.6 Cycle operation
5.5.5.7 Retrieval of the handpiece and BI
5.5.5.8 Spore recovery and incubation
5.5.5.9 Acceptance criteria
27 5.6 Mechanical air removal
5.6.1 Air removal (dynamic-air-removal sterilizers)
5.6.1.1.1 Bowie-Dick-type test pack
5.6.1.1.2 Test pack placement
5.6.1.1.3 Cycle operation
5.6.1.1.4 Acceptance criteria
5.6.2 Air leaks (prevacuum sterilizers)
5.6.2.1 General
5.6.2.2 Test procedure
28 5.6.2.3 Acceptance criteria
5.7 Moisture retention
5.7.1 Textile test packs
5.7.1.1 Test pack construction
5.7.1.2 Pack placement
5.7.1.3 Cycle operation
5.7.1.4 Out-of-chamber handling
29 Figure 4—Folding a surgical towel to yield nine plies (see 5.7.1.1)
5.7.1.5 Acceptance criteria
5.7.2 Wrapped instrument test trays
5.7.2.1 Test tray composition
5.7.2.2 Tray placement
5.7.2.3 Cycle operation
5.7.2.4 Out-of-chamber cooling
30 5.7.3 Paper–plastic peel pouches
5.7.3.1 Pouch preparation
5.7.3.2 Pouch placement
5.7.3.3 Cycle operation
5.7.3.4 Out-of-chamber cooling
5.7.3.5 Evaluation of pouch for moisture retention
5.7.2.5 Evaluation of tray for moisture retention
5.8 Sterilizer performance certification and recordkeeping
31 Annex A (informative) Rationale for the development and provisions of this standard
36 Annex B (informative) Examples of temperature during the holding time
Figure B.1—Unacceptable process
Figure B.2—Unacceptable process
37 Figure B.3—Unacceptable process
Figure B.4—Unacceptable process
Figure B.5—Acceptable process
38 Figure B.6—Acceptable process
39 Annex C (informative) Examples of cycle documentation
Figure C.1—Schematic of cycle phases
40 Figure C.2—Example of a cycle printout
41 Bibliography
42 For further reading
AAMI ST55 2016 R2023
$125.88